What you will learn
- One atom is too light to weigh, so masses are compared with 1/12 of a carbon-12 atom. A ²⁴Mg atom has a relative isotopic mass of 24 (no unit).
- Magnesium is a mixture of three isotopes. We need the relative mass of each and how common it is (its abundance). A mass spectrometer measures both.
- A time of flight mass spectrometer has four stages in a vacuum: ionisation, acceleration, ion drift and detection.
- Electron impact: a high-energy electron knocks an electron out of each gaseous atom. Mg(g) + e⁻ → Mg⁺(g) + 2e⁻
- A negatively charged plate attracts the positive ions; the electric field gives every ion the same kinetic energy.
- With the same kinetic energy, KE = ½mv² means a lighter ion moves faster, so it takes less time, t = d ÷ v, and arrives first.
- At the negatively charged detector each ion gains an electron. The electrons flowing make a current; the bigger the current, the more ions.
- Each peak's m/z is an isotope's relative mass (for 1+ ions); its height is the relative abundance; the pattern identifies the element. Name each peak by its ion, such as ²⁵Mg⁺.
- Electrospray adds a proton without breaking the molecule: M(g) + H⁺ → MH⁺(g). So Mr is 1 less than the peak: aspirin's peak at 181 gives Mr 180.
- Mass of one ion = M ÷ 1000 ÷ L (in kg); v = √(2KE ÷ m); t = d ÷ v. The distance is the same for every ion, so t is proportional to √m.
Notes
- Relative isotopic mass: the mass of an atom of an isotope compared with 1/12 of the mass of an atom of carbon-12. No unit.
- Abundance: how common an isotope is, as a percentage of the element's atoms.
- A time of flight (TOF) mass spectrometer, in a vacuum:
- Ionisation. Electron impact: high-energy electrons from an electron gun knock an electron out of each gaseous atom, X(g) + e⁻ → X⁺(g) + 2e⁻. Electrospray (larger molecules): the sample, dissolved in a volatile solvent, is sprayed from a fine needle at a high positive voltage; each molecule gains a proton, M(g) + H⁺ → MH⁺(g).
- Acceleration. A negatively charged plate attracts the positive ions: an electric field gives every ion the same kinetic energy.
- Ion drift. No field in the flight tube. KE = ½mv², so lighter ions move faster and arrive first; t = d ÷ v.
- Detection. Each ion gains an electron at the negatively charged detector; the electrons flowing make a current, bigger when more ions arrive.
- The spectrum: relative abundance against m/z. For 1+ ions m/z is the relative isotopic mass. Name a peak by its ion (²⁵Mg⁺, not ²⁵Mg).
- Electrospray: the peak is MH⁺, so Mr = peak − 1.
- Time of flight: mass of one ion in kg = molar mass in g mol⁻¹ ÷ 1000 ÷ 6.022 × 10²³; v = √(2KE ÷ m); t = d ÷ v.
About this model
Every particle is labelled: e⁻ electron, H⁺ proton, ions and atoms with their symbols. Atoms and ions are drawn far larger than real. In the moving pictures the differences in speed are exaggerated so that the ions separate on screen; in a real instrument ²⁴Mg⁺ arrives only about 4% sooner than ²⁶Mg⁺, as the numbers in Try it show.
Isotope masses are taken as whole numbers (24, 25, 26), as in exam questions; measured values differ slightly (²⁴Mg 23.985). Abundances: ²⁴Mg 79.0%, ²⁵Mg 10.0%, ²⁶Mg 11.0%. The average, relative atomic mass, is defined and calculated in the next lesson.
Exam Prep is laid out as on an AS paper; the questions are written for this site. The periodic table has the relative atomic masses of the AQA data booklet. Values are listed in the topic's values file.